Operating principle 
      
  
  
    
      
      
Interface measurement method - 1st echo
      
  
  
  
   
            
    
        
      
Measurement range
      
  
  
    
      
      
Varies according to probe and material, 0.60 mm to 508 mm in interface - 1st echo measurement mode, 2.00 mm to 127.0 mm in multi-echo measurement mode, the coating thickness range will be 0.3 mm to 2.50 mm.
      
  
  
  
   
            
    
        
      
Measurement resolution
      
  
  
    
      
      
Default 0.01 mm - selectable 0.01, 0.1 mm.
      
  
  
  
   
            
    
        
      
Material celerity range
      
  
  
    
      
      
0,508 à 18,699 mm/ms 
      
  
  
  
   
            
    
        
      
Resolution of the celerity of the material 
      
  
  
    
      
      
1 m/s
      
  
  
  
   
            
    
        
      
Units 
      
  
  
    
      
      
Pouces ou millimètres
      
  
  
  
   
            
    
        
      
Calibration
      
  
  
    
      
      
 One-point, on-block and off-block calibration, two-point calibration
      
  
  
  
   
            
    
        
      
Excitation pulse generator 
      
  
  
    
      
      
Voltage
      
  
  
  
   
            
    
        
      
Transient pulse generator 
      
  
  
    
      
      
120 V at 50 ohms, with 20 MHz oscilloscope
      
  
  
  
   
            
    
        
      
Receiver bandwidth 
      
  
  
    
      
      
500 KHz à 12 MHz @ -3 dB
      
  
  
  
   
            
    
        
      
Receiver gain 
      
  
  
    
      
      
Automatic gain control
      
  
  
  
   
            
    
        
      
Display type 
      
  
  
    
      
      
High-resolution graphic liquid crystal display, 64 x 128 pixels, 53.0 mm x 27.0 mm with backlight and contrast adjustment
      
  
  
  
   
            
    
        
      
Recurrence frequency 
      
  
  
    
      
      
4 Hz or 8 Hz, user selectable, 24 Hz acquisition mode capture rate
      
  
  
  
   
            
    
        
      
Display of thickness value 
      
  
  
    
      
      
NORMAL display mode 5 digits, 10.6 mm high
      
  
  
  
   
            
    
        
      
Display mode 
      
  
  
    
      
      
B-SCAN 5 digits, 2.55 mm high.
      
  
  
  
   
            
    
        
      
Solid or hollow figures indicate a coupled or decoupled condition.
      
  
  
    
      
      
Display of last playback 
      
  
  
  
   
            
    
        
      
Configurations 
      
  
  
    
      
      
9 standard probe configurations.
      
  
  
  
   
            
    
        
      
Setting alarms
      
  
  
    
      
      
Minimum and maximum alarms, range 0.25 mm to 508 mm, value alternates between full and empty when alarms are activated and violated
      
  
  
  
   
            
    
        
      
Power supply requirements 
      
  
  
    
      
      
2 AA batteries
      
  
  
  
   
            
    
        
      
Battery life / Autonomy 
      
  
  
    
      
      
Approximately 60 hours
      
  
  
  
   
            
    
        
      
Switching off the instrument 
      
  
  
    
      
      
Selectable - ALWAYS ON or AUTO OFF after 5, 10, 15, 30 minutes of inactivity
      
  
  
  
   
            
    
        
      
Language 
      
  
  
    
      
      
Selectable - English, German, French, Spanish, Italian, Russian, Japanese and Chinese
      
  
  
  
   
            
    
        
      
I/O connectors 
      
  
  
    
      
      
Dual LEMO 00 probe (coaxial)
      
  
  
  
   
            
    
        
      
USB connector 
      
  
  
    
      
      
Mini-USB vers PC
      
  
  
  
   
            
    
        
      
Operating temperature 
      
  
  
    
      
      
-10° C à +50° C
      
  
  
  
   
            
    
        
      
Storage temperature 
      
  
  
    
      
      
-20° C à +60° C
      
  
  
  
   
            
    
        
      
Weight 
      
  
  
    
      
      
 223 g (including batteries)
      
  
  
  
   
            
    
        
      
Dimensions 
      
  
  
    
      
      
138 mm x 32 mm x 75 mm
      
  
  
  
   
            
    
        
      
Shocks 
      
  
  
    
      
      
IEC 68-2-27 Ea, according to Mil Std 810C Method 516.2 Procedure I with a 15 g 11 ms semi-sinusoidal wave pulse applied 6 times per axis
      
  
  
  
   
            
    
        
      
Waterproofing 
      
  
  
    
      
      
IEC529 / IP54, dust and drip-proof to IEC 529 specifications for IP54 classification.